Loading of dental implants immediately after placement is an important attribute for the patients receiving the implants, but may induce a failure in the osseointegration of the implant. Even though the loading required to induce such failures is currently unknown, it is estimated that the osseointegration may fail if an implant is luxated in bone by more than . Therefore, the ability to measure this loading can provide the practicing dentist with critical information in estimating the functional life of a newly placed implant. This paper discusses the design and fabrication of a cost-effective test setup for measuring the amount of horizontal force required to displace a nonosseointegrated implant, and presents the results of our pilot in vitro load measurements for dental implants mounted in a bovine mandible. Although the sample size of our measurements is not statistically significant, the initial data show that the amount of horizontal force required to displace a implant by is in the order of , and therefore the implant may fail to osseointegrate for biting forces that are as low as , which is about half of the typical biting force of an adult in the molar area. One implication of our study is that implants having smaller diameters may move and fail to osseointegrate for even lower biting forces. Furthermore, our work represents the first steps in developing appropriate metrics to correlate the measured biting force generated by a patient with his or her candidacy for immediate functional implant loading.
Skip Nav Destination
e-mail: ilies@engr.uconn.edu
Article navigation
March 2008
Technical Briefs
Measuring Luxation of Dental Implants In Vitro
Horea T. Ilieş,
Horea T. Ilieş
Department of Mechanical Engineering,
e-mail: ilies@engr.uconn.edu
University of Connecticut
, Storrs, CT 06269
Search for other works by this author on:
Matthew Raby,
Matthew Raby
Department of Mechanical Engineering,
University of Connecticut
, Storrs, CT 06269
Search for other works by this author on:
Richard Stevenson
Richard Stevenson
Department of Mechanical Engineering,
University of Connecticut
, Storrs, CT 06269
Search for other works by this author on:
Horea T. Ilieş
Department of Mechanical Engineering,
University of Connecticut
, Storrs, CT 06269e-mail: ilies@engr.uconn.edu
Dennis Flanagan
Matthew Raby
Department of Mechanical Engineering,
University of Connecticut
, Storrs, CT 06269
Richard Stevenson
Department of Mechanical Engineering,
University of Connecticut
, Storrs, CT 06269J. Med. Devices. Mar 2008, 2(1): 014501 (4 pages)
Published Online: March 7, 2008
Article history
Received:
May 30, 2007
Revised:
January 2, 2008
Published:
March 7, 2008
Citation
Ilieş, H. T., Flanagan, D., Raby, M., and Stevenson, R. (March 7, 2008). "Measuring Luxation of Dental Implants In Vitro." ASME. J. Med. Devices. March 2008; 2(1): 014501. https://doi.org/10.1115/1.2889057
Download citation file:
Get Email Alerts
Cited By
Development and prospects of flexible puncture needle: a review
J. Med. Devices
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
Related Articles
Design and Implementation of Noncontact Measurement Device for Irregular Dental Osseointegration Detection
J. Med. Devices (June,2011)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Design and Validation of a Machine for Reproducible Precision Insertion of Femoral Hip Prostheses for Preclinical Testing
J Biomech Eng (April,2000)
Determining the Fatigue Life of Dental Implants
J. Med. Devices (March,2008)
Related Proceedings Papers
Related Chapters
Novel and Efficient Mathematical and Computational Methods for the Analysis and Architecting of Ultralight Cellular Materials and their Macrostructural Responses
Advances in Computers and Information in Engineering Research, Volume 2
Application of Non-Linear Elastic Wave Spectroscopy (NEWS) to In Vitro Damage Assessment in Cortical Bone
Biomedical Applications of Vibration and Acoustics in Imaging and Characterizations
Computer Aided Oracle Bone Inscriptions Textual Research Based on Ontology
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3